MIL-PRF-49446B(CR)
shall be used in responsivity calculations and blackbody temperature shall be 500°, ±5oK.
Failure to meet requirements of 3.5.7 shall constitute failure of this test.
4.5.8 Temperature sensors. The 77o K calibration point may be determined by cooling
the inner Dewar with liquid nitrogen. A 1.00 mA ±0.01 mA bias shall be applied to the sensor
diode and the voltage shall be monitored and recorded at the 77o K calibration point. Failure to
meet requirements of 3.5.8 shall constitute failure of this test.
4.5.9 Spectral response. The spectral response of the sixth element from each end of the
array or adjacent element on the high end and the higher number of the two center elements shall
be tested per "Standard Procedures for Testing Infrared Detectors and Describing Their
Performance", by R.C. Jones, D. Goodwin, and G. Pullen, dated 12 September 1960, Office of
the Director of Defense Research and Engineering. If these elements are defective an adjacent
element may be substituted. The spectral sensitivity is defined as the 50 percent relative response
points and shall be measured on the assembled Detector-Dewar. Failure to meet requirements of
3.5.9 shall constitute failure of this test.
4.5.10 Detector bias. The detector bias power shall be tested per "Standard Procedures
for Testing Infrared Detectors and Describing Their Performance", by R.C. Jones, D. Goodwin,
and G. Pullen, dated 12 September 1960, Office of the Director of Defense Research and
Engineering. Failure to meet requirements of 3.5.10 shall constitute failure of this test.
4.5.11 Dewar warm and cold shields. During in-process inspection, the warm and cold
shield shall be measured to ensure conformance with the requirement Alignment of the cold
shield relative to detector elements shall be checked 100 percent during in-process inspection.
Failure to meet requirements of 3.5.11 shall constitute failure of this test.
4.5.12 Heat load. The heat load of the Detector-Dewar shall be determined with all
detectors operating at specified bias. The heat load of the Detector-Dewar shall be determined by
partially filling the Dewar's inner stem with liquid nitrogen and monitoring the rate of loss of the
cryogen. The heat load shall be calculated from the average or integrated loss rate of the cryogen
for the last 5 minutes of the Dewar's hold-time. For calculations, cooling produced by the
nitrogen phase change only shall be considered as total useful energy. If this test is performed
with the detector array unbiased, the results of this test shall be added to the results of the
detector bias power test to determine the heat load. Failure to meet requirements of 3.5.12 shall
constitute failure of this test.
4.5.13 Vacuum integrity. Vacuum integrity shall be determined on a basis of heat load
capacity. Heat load shall be measured on the unbiased Detector-Dewar. The temperature of the
Detector-Dewar shall then be elevated to +71°C for a period of 72 hours. At the completion of
the temperature exposure and stabilization at ambient temperature, the heat load shall be
measured again. Getters may not be fired during this test. Failure of the Detector-Dewar to meet
the requirements of 3.5.12 (heat load) or a 15 percent, or greater, increase in heat load between
the measurement before and after temperature exposure shall constitute failure of this test.
11
For Parts Inquires call Parts Hangar, Inc (727) 493-0744
© Copyright 2015 Integrated Publishing, Inc.
A Service Disabled Veteran Owned Small Business